Care and Cleaning
EN
The following processes is recommended, but should be adapted to comply with local
institution protocols. If you are uncertain, you should seek advice from your local Infection
Control Specialist.
The AtmosAir MRS should be routinely decontaminated between patients and at regular
intervals while in use.
Fire Barrier
recommended that Personal Protective Equipment (PPE) is used during handling of the
Fire barrier should be replaced if soiled or damaged.
Cover Cleaning Options
Sewn AtmosAir
1.
Remove or push bed linens to center of mattress.
2.
Wipe and rinse any soiling from the mattress surface and base.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
After wipe down, dry surface with towel.
5.
6.
Clean pump and tubes (if needed) by wiping with a damp cloth.
RF Welded AtmosAir
1.
Push bed linens to center of mattress to wipe down, remove to launder top cover
or wipedown as described below.
2.
Wipe and rinse any soiling from the mattress surface and base.
Use 1000 ppm chlorine or 70% alcohol.
3.
Rinse with a wet towel to remove chemicals.
4.
After wipe down, dry surface with towel.
5.
8
Do not use Phenol-based solutions or abrasive componds or pads
on the cover during decontamination process as these will damage
the surface coating. Do not boil or autoclave the cover.
Covers should be wiped down immediately after it has been exposed
The sewn AtmosAir MRS cover is cleaned by wipe down methods only.
Do not launder as damage to the base may occur.
The RF Welded AtmosAir MRS has a detachable top cover that can be
laundered. However the base is cleaned by wipe down methods only.
Do not launder base as damage may occur.