Atmosair Pump Installation; Led Indicators - arjo AtmosAir 9000A Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 74

AtmosAir Pump Installation

Pump installation
1.
Place pump on a solid stationary surface or suspend on end of bed frame with
built-in hanger.
2.
Attach pump hoses to mattress, as shown with (Figure 1) below:
• Attach hoses to the mattress connectors on the patient left side of
the mattress for alternating p ressure.
• Ensure pump hoses are attached to pump.
3.
Ensure air hoses are not kinked and will not be pinched by any articulated bed
mechanisms.
4.
Attach hose connector to the pump.
5.
Plug pump unit into a properly-grounded wall outlet.
6.
Verify power to this outlet is not controlled by a wall switch.
7.
Switch pump on.

LED Indicators

1.
Green LED is lit when the pressure is at the set level.
2.
Yellow LED is lit when the pressure is below the set level. Check connections.
MAX.
ALT. PRESSURE / ROTATION
6
Mattress
Connectors
Tag
Figure 1 Mattress Hose Connector
MIN.
NORMAL PRESSURE
LOW PRESSURE
POWER
Figure 2 Control Panel
Pump Hoses
ON
OFF
AtmosAir