Télécharger Imprimer la page

Invacare Jasmine Patient Lift Manuel D'utilisation Et D'entretien page 24

Publicité

Les langues disponibles

Les langues disponibles

SECTION 7—MAINTENANCE
Replacing the Leg Actuators
NOTE: For this procedure, refer to FIGURE 7.3.
NOTE: If the legs of the patient lift do not open and close properly, it may be necessary to replace one or both of the leg actuators. 
1. Test the operation of the legs:
Press the close button on the hand control to close the legs.
Press the open button on the hand control to open the legs.
2. While holding on to the mast handles, carefully tip the patient lift back until it rests on the floor.
NOTE: It is necessary to remove both small base covers before removing the large base cover.
3. Remove the sixteen screws that attach the small base covers to the base.
4. Remove the eight pan head screws that attach the large base cover to the base.
NOTE: If the left or the right leg of the patient lift does not operate properly, perform the following steps on the side of the base necessary 
to replace the actuator. If both legs do not operate properly, perform this replacement procedure on both sides of the base.
5. Locate and disconnect the leg actuator lead wires. 
NOTE: The actuator lead wires will be visible after the large base cover is removed.
6. Remove the socket head screw and washer that attach the leg actuator to the base.
7. At the other end of the actuator, remove the socket head screw and nut that attach the leg actuator to the leg bracket.
8. If necessary, repeat STEPS 5‐7 on the opposite side of the base to remove the other leg actuator.
9. To replace the actuator(s), reverse STEPS 5‐7.
Pan head screw
(eight screws attach the
large base cover)
Jasmine™ Patient Lift
All manuals and user guides at all-guides.com
Caster
Leg
Large Base
Cover
FIGURE 7.3 Replacing the Leg Actuators
BOTTOM VIEW OF
PATIENT LIFT
Mast Handles
24
Screw
(eight screws attach each
small base cover)
Small Base Cover
Part No 1150704

Publicité

loading