SECTION 6—TROUBLESHOOTING
SECTION 6—TROUBLESHOOTING
SYMPTOMS
Patient Lift feels loose.
Noisy or dry sound from pivots.
Electric actuator fails to lift when button
is pressed.
Unusual noise from actuator.
Lift arms will not lower in uppermost
position.
Lift arms will not lower during a power
retraction.
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
Jasmine™ Patient Lift
All manuals and user guides at all-guides.com
FAULTS
Mast/Base joint loose.
Tie - Rods are loose.
Needs lubrication.
Hand-control or actuator connector
loose.
Battery low.
RED emergency stop button pressed IN.
Battery not connected properly to con-
trol box.
The connecting terminals are damaged.
Electric actuator in need of service or
load is too high.
Actuator is worn or damaged or spindle
is bent.
Lift arms require a minimum weight load
to lower from the uppermost position.
Shoulder bolt at the junction of the boom
and mast may not be properly installed.
20
SOLUTION
Refer to Replace the locknut. on page 26.
Refer to Lubricating the Lift on page 22.
Charge batteries. Refer to Charging the
Battery on page 13.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Charging the Battery on page 13.
Replace the battery pack. Refer to Charg-
ing the Battery on page 13.
Refer to Replacing the Boom Actuator on
page 22. Contact your Dealer.
Refer to Replacing the Boom Actuator on
page 22.
Pull down slightly on the lift arms.
Refer to Replacing the Swivel Bar or
Optional Cradle Attachment on page 25.
Part No 1150704