Toggle mode generates a repetitive pulse stream that toggles between two load levels. It causes the load input to
alternate between two pre-defined levels as in continuous operation except that the transient points are controlled
by explicit triggers instead of an internal transient generator.
Step 1– Configure the sequence properties
Configure the Sequencer Toggle accordingly. Refer to the below table for details.
Parameter
Available Key Settings
Transient Setting
minimum to maximum
(I
, V
, P
, or R
)
1
1
1
1
Current Mode
Fix, Stp, or Lst
Voltage Mode
Resistance Mode
Power Mode
Trigger Source
Key, IO, or Rmt
Trigger Delay
0 to 0.255 s
Keysight EL30000 Series User's Guide
Sets the transient voltage, current, resistance, or power level in voltage, ampere, ohms, or
watts.
Sets the current/voltage/resistance/power mode. This determines what happens to the input
current/voltage/resistance/power when the system is initiated or triggered.
Default is Fix.
Action required:
Press Curr Mode, Volt Mode, Res Mode, or Pow Mode to toggle between Fix, Stp, and Lst.
Fix (Fixed)
Keeps the input at its immediate value.
Stp (Step)
Steps the input to the triggered level when a trigger occurs
Lst (List)
Causes the input to follow the list values when a trigger occurs.
Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the trigger delay in seconds.
Default is 0 s.
Description
Selects the Run Stopped softkey as a trigger source.
Selects any configured digital IO (with Trigger Input
function) as a trigger source
Selects a remote interface command as a trigger
source.