From the remote interface:
To configure the load operating mode to CV mode:
FUNC VOLT, (@1)
To set the sequencer to Pulse mode:
TRAN:MODE PULS, (@1)
To configure main voltage level with 5 V:
VOLT 5, (@1)
To configure the value for the transient voltage level with 10 V:
VOLT:TLEV 10, (@1)
To configure the transient pulse width with 0.5 s :
TRAN:TWID 0.5, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
VOLT:MODE LIST
To enable input 1:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)