From the remote interface:
To configure the load operating mode to CC mode:
FUNC CURR, (@1)
To set the sequencer to LIST mode:
TRAN:MODE LIST, (@1)
To configure the List current of input 1 with 0.1 A, 0.2 V, 0.3 A, 0.4 A and 0.5 A:
LIST:CURR 0.1,0.2,0.3,0.4,0.5, (@1)
To configure all the List time of input 1 with 1 s:
LIST:DWELL 1,1,1,1,1, (@1)
To enable all the List BOST of input 1:
LIST:TOUT:BOST 1,1,1,1,1, (@1)
To disable all the List EOST of input 1:
LIST:TOUT:EOST 0,0,0,0,0, (@1)
To set the current mode of input 1 to List:
CURR:MODE LIST, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:SOUR IMM, (@1)
To enable input 1:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)
92
Keysight EL30000 Series User's Guide

Publicité

loading

Ce manuel est également adapté pour:

El33133aEl34143aEl34243a