From the remote interface:
To configure the load operating mode to CV mode:
FUNC VOLT, (@1)
To set the sequencer to Continuous mode:
TRAN:MODE CONT, (@1)
To configure main voltage level with 5 V:
VOLT 5, (@1)
To configure the value for the transient voltage level with 10 V:
VOLT:TLEV 10, (@1)
To configure the transient frequency with 50 Hz :
TRAN:FREQ 50, (@1)
To configure the transient duty cycle with 10.5%:
TRAN:DCYC 10.5, (@1)
To configure trigger delay with 0.2 s:
TRIG:TRAN:DEL 0.2, (@1)
To configure the transient repeat count with 20:
TRAN:COUN 20, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
VOLT:MODE LIST
To enable input 1:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)
Keysight EL30000 Series User's Guide
95