Installing the optional GPIB interface (EL34143A and EL34243A only)
TURN OFF POWER AND REMOVE ALL CONNECTIONS PRIOR TO INSTALLATION
Turn off the power and remove all connections, including the power cord, from the instrument prior install-
ation of the GPIB interface.
RETAIN GPIB COVER PLATE
After installing the GPIB option, retain the cover plate for use in the event that you ever remove the GPIB
option. The instrument should never be connected to AC mains or input terminal when the rear-panel opening
is not securely covered with either GPIB module or the cover plate.
The following tools are required.
– Torx screwdriver
Make sure you turn the electronic load upside down before proceeding.
Use a Torx screwdriver to remove the M3 screws
from the GPIB cover plate. Retain the screws for
use later in this procedure.
Then, remove the cover plate.
Attach the cable to the connector located in the
previous step.
Keysight EL30000 Series User's Guide
Locate the connector, just inside the unit, to connect
the cable that is attached to the GPIB module.
Put the module into the unit. Use the screws that
was removed earlier to secure the GPIB plate in
place.
This concludes the GPIB installation procedure.