– Leave the default Duration and Sample Period at 30 s and 200 ms respectively.
– Set the Trigger Source to Run/Stop Key. Press Trigger Settings > Trigger Source > Run Key .
Step 6 – Configure and enable the DUT output accordingly
Step 7 – Turn on Input 1, start the sequencing and log the data
Press Back to return the display to the data logger view.
– Press Input 1 [On] to turn on Input 1.
– Press Datalog Run to run the data logger. The data logger will be initiated and Input 1 trace is displayed on the
screen.
– Press Sequencer > Sequencer LIST > Run to start input sequencing. Alternatively, press [List Run/Stop] .
The data logger will run for 30 seconds and log the voltage data from input 1. When the data log completes, press
Auto Scale . You should see the input waveform displayed as follows:
A message will indicate that the logged data has been saved in a file named default.dlog. If you wish to save the data
under a different filename, you must specify the filename before the data logger runs.
Press File Name located at the data logger Properties menu to specify a filename.
Step 8 – Export the data
After you have completed the data logging, you can use the Export File key to export the logged data to a .csv
(comma separated values) file.
Keysight EL30000 Series User's Guide
103