Step 2 – Run the input sequence
In this example, the load operation is in CC mode.
– Set I
, trigger delay and repeat count accordingly: Use the navigation keys to select the field; use the numeric
1
entry keys to enter the value. The value is set when you press [Enter] .
– Set Trigger Source to Key. Press Trig Src Key .
– Press the color-coded [On] key to enable the selected input.
– Press Run to start the sequence operation. To abort the operation, press Stop .
– Press Back to exit and return to the previous menu.
From the remote interface:
To configure the load operating mode to CC mode:
FUNC CURR, (@1)
To set the sequencer to Toggle mode:
TRAN:MODE TOGG, (@1)
To configure the main current level with 5 A:
CURR 5, (@1)
To configure the value for the transient current level with 10 A:
CURR:TLEV 10, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
CURR:MODE LIST
To enable input 1:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)
100
Keysight EL30000 Series User's Guide