Maintenance
9
Maintenance
The uni-cart, pro-cart, duo-cart, compact-cart, classic-cart or endo-cart has been developed and
constructed for many years of trouble-free use. Check the functional capability of the following
parts every 12 months in order to guarantee safety.
•
Monitor shelf:
Swivels and tilts easily without too much play.
•
Shelves:
Check whether the mounting screws have been tightened and whether the shelf is stable
and flat.
•
Castors:
•
Check that the castors run freely, the brakes operate correctly.
Check that the 4 bolts holding the castors to the bottom side of the base and the castors
•
themselves are seated firmly in their holders.
The running surfaces of castors must be free from contamination, so that they can function
correctly.
•
Socket strips:
Check the main cable for damage and firm seating.
•
Auxiliary sockets:
Check the cable for damage and firm seating.
•
Variabel height support arms (flexion-port):
The height adjustment functions freely, the raising force is adjusted to the weight of the
device.
•
Support arms:
Rotating and tilting functions smoothly, without excess play.
•
Isolating transformers:
Safety-relevant checks of isolating transformers.
•
Serial number:
Compare the serial number of the cart with the data of the equipment log book.
•
Fuses:
Check whether the correct fuses have been installed.
If you encounter any problems during these checks you should contact your supplier immediately.
48