Electrical safety
3.7
Cable connection
Users of the uni-cart, pro-cart, duo-cart, compact-cart, classic-cart or endo-cart carrier system
without an isolating transformer should ensure that the cable connection between the cable
system of the cart and the equipment is a permanent connection and can only be removed using
tools. The plug pull-out lock can be installed or removed by two screws, using a flat screwdriver
or Allen key.
3.8
Combination of equipment
The following should be observed for combination of equipment on the carrier system:
•
Auxiliary equipment connected to analogue and digital interfaces of the equipment must
be certified in compliance with the relevant EN specifications (e.g. EN 60950 for data-
processing equipment and EN 60601 for medical electrical equipment).
•
Furthermore, all configurations must comply with the valid version of the system standard
EN 60601-1-1. Anyone connecting additional equipment to the signal input or signal output
is a system configurer and is therefore responsible for ensuring compliance with the valid
version of the system standard EN 60601-1-1.
If you have any questions, you should contact your local dealer or technical services.
Please note:
Make sure that this is also the case for adaptation of equipment in the power supply circuit
(e.g. multiple socket strip).
3.9
EMC
Electromagnetic compatibility of the medical electrical equipment positioned on the cart is only
checked for the combination supplied by the overall system configurer. If you use a different
equipment combination, you should check electromagnetic compatibility between individual
items of equipment before using in a medical setting.
The carrier system cannot be used in an NMR environment due to the presence of ferro-
magnetic materials.
ITD GmbH excludes any liability in this respect.
40