Commissioning
65
P51*00 – Connection to a Shunt Resistor
K
H
L
For measurements in the mV range, twist the measuring cables (H)
and connect them to the corresponding terminals (K) of the shunt.
Twist the measuring cables until close to the housing.
Fasten the housing on a plane mounting surface using four M6 screws,
each with a washer (L). Tighten the screws to the recommended torque of
2 Nm.