104
K0320 Issue No. 1
0-20 mA
4-20 mA
mA out
Control system
Regelsystem
Système de régulation
4-20 mA
XMTR
Control system
Regelsystem
Système de régulation
Temperature transmitter
simulation
User can dial a temperature in
degrees to output a 4-20 mA or 0-20
mA signal. This signal can be used to
simulate transmitters at control system
input terminals. The simulated
transmitter may be linearized or non-
linearized. Function is available for all
type of temperature sensors supplied
as standard in the MCX-II.

Setting up a simulation

Leave the cursor on NONE in the input
menu and select the required RTD or
T/C function (ignore the
CJ compensation menu). In the mode
screen, select SIM to simulate a
transmitter and proceed to select all
transmitter characteristics. Fill-out the
transmitter span and press EXE to
open the working screen. The lower
windows shows the MCX-II output
milliamps in small digits.