Special Output Functions

Ramp Mode

The MCX-II can generate automatically
a time linear increasing or decreasing
output signal. Signal limits are free to
set within any of the available output
ranges. Time scales can be set
between 1 and 9999 seconds.
NOTE: The automatic ramp mode
cannot be used together with the
frequency measurement mode and
the pulse count mode.
IN P U T
O H M
F R E Q .
C O U NTE R
S WITC H
B A TTE R Y

Setting up the ramp mode

Select RAMP and fill out the set-up
screen. The high-end value sets the
maximum output. The low-end value
sets the minimum output. Fill out the
required travel and dwell time in
seconds and press EXE.
O U TP U T
mA/ XMT
mV
V O L T
T/ C
R TD