82
K0320 Issue No. 1
Output of electrical signals
Ohms
Go through the OHM menu and select
either the 0-400 or the 0-2000 Ohms
range and enter the DIRECT mode.
Resistance is simulated across the two
terminals as shown. Connecting a third
or fourth wire adapts the simulation to
3 or 4 wire input. Adjustment
resolution is respectively 0.01 Ohms
and 0.1 Ohms.
Frequency
Pulses are symmetrical square wave
and -70 mV zero based. Frequency is
set by the
and
key-pad. Resolution is 0.1 Volt and
amplitude is default set to 5.0 Volts.
With frequency set at 0 Hz this
function becomes available as a
voltage adjustable and regulated DC
supply. Frequency output can take
loads up to 34 mA at 24 Volts
maximum.
Frequency range 0-100 Hz
Go to pulse DIRECT mode to output any
pulse signal between zero and 100 Hz.
Adjustments are made with 0.01 Hz
resolution.
keys or numerical