GE Druck MCX-II Guide De L'utilisateur page 120

Multi-calibrateur
Table des Matières

Publicité

116
K0320 Issue No. 1
Service, repair and parts
NOTES:
1. Calibration standards should
have a valid calibration
certification and should be at
least 3 times more accurate as
the published accuracies for
the MCX-II for each range.
2. As derived from electrical
parameters, temperature
sensor ranges do not require
any individual recalibration.
3. Be sure that you use the test
leads, as standard supplied
with the MCX-II or other low
e.m.f. test leads.
4. The MCX-II should be powered
from the on board batteries
during recalibration.
5. Recalibration must be carried
out at reference ambient
temperature and humidity.
MCX-II must be exposed to this
condition for at least 2 hours
before the actual recalibration
starts.
6. Recalibrate mV ranges before
starting recalibration of the
Cold Junction Compensation
system.
7. When accessing the calibration
menu the temperature unit is
set to °C and the temperature
scale is set to IPTS 68.
8. Calibration is recommended
between 12 and 14 months.