28
K0320 Issue No. 1
Power Sources
You may charge for periods longer
than 14 hours. Note that at lower
ambient temperatures the capacity of
Ni-Cd batteries is significantly lower. If
the batteries do not reach their normal
capacity after a 14 hours charging
period, cycle complete discharging
and charging for at least 2 times. If
batteries remain weak they should be
replaced. No particular brand of Ni-Cd
is recommended although cells rated
at 2.0 Ah have preference over
general available 1.8 Ah cells.
Operating from 115 V or 230V line voltage
(50 or 60 Hz)
Use the Line Adaptor/Charger #13603
only; other equipment may cause
damage the calibrator. Check the
indicated line voltage. Connect the
appropriate Line Adaptor/Charger
plugs to the line and to the calibrator.
The green led EXT at the calibrator will
now be on.