Changing From Alkaline To Rechargeable Batteries; Recharging - GE Druck MCX-II Guide De L'utilisateur

Multi-calibrateur
Table des Matières

Publicité

26
K0320 Issue No. 1
Power Sources
Changing from Alkaline to
Rechargeable batteries
Remove the Alkaline batteries.
Remove the switch protector sticker
and slide the switch to the "Ni-Cd"
position.
Place a new protector sticker to save
the new switch position. Install 6
rechargeable batteries (purchased
locally). Battery switch position will be
confirmed on the screen during start-
up of the calibrator. Recharge
batteries when the screen shows the
blinking battery symbol.
CAUTION:
Do not fit alkaline or other non-
rechargeable batteries after you
have changed the switch position to
"Ni-Cd".

Recharging

Use the Line Adaptor/Charger #13603
only; other equipment may cause
damage to the calibrator. Check the
indicated line volt switch in the
CHARGE position. Connect the
appropriate Adaptor/Charger plugs to
the line and to the calibrator. The
amber led CHG at the calibrator power
panel will now be on. As adaptor and
charger functions are fully
independent from each other, the
calibrator can be used while charging.
Recharging time from complete
discharge to fully charge is 14 hours.

Publicité

Table des Matières
loading

Table des Matières