Endress+Hauser RIA452 Manuel De Mise En Service page 87

Masquer les pouces Voir aussi pour RIA452:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RIA452
Installation instructions
Environment
Endress+Hauser
Resolution
Temperature drift
10.0.5
Installation
Mounting location
Panel, cutout 92x92 mm (3.62x3.62") (see 'Mechanical construction').
Orientation
Horizontal +/- 45 in every direction
Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
Storage temperature
-30 to +70 °C (-22 to 158 °F)
Operating height
< 3000 m (9842 ft) above MSL
Climate class
As per IEC 60654-1, Class B2
Condensation
Front: permitted
Casing: not permitted
Degree of protection
Front IP 65 / NEMA 4
Casing IP 20
Shock and vibration resistance
2(+3/-0) Hz - 13.2 Hz: ±1.0 mm
13.2 Hz - 100 Hz: 0.7 g
Electromagnetic compatibility (EMC)
• Interference immunity:
To IEC 61326 industrial environments / NAMUR NE 21
• Interference emissions:
To IEC 61326 Class A
13 bit
0.1%/10K (0.056%/10 °F)
Technical data
87