Occlusion Pressure Limits For Ivac 50Ml Syringes - IVAC P5000 Mk II Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Specifications

Occlusion Pressure Limits for IVAC 50ml Syringes

The following tables show the worst case values for line pressure, time to alarm and bolus volume
that can be expected in the event of an occlusion when the IVAC 50ml syringe is selected,
G40020 administration set.
Alarm Level
Rate (ml/h)
0
1.0
1
1.0
2
1.0
3
1.0
4
1.0
5
1.0
6
1.0
7
1.0
8
1.0
9
1.0
10
1.0
0
5.0
1
5.0
2
5.0
3
5.0
4
5.0
5
5.0
6
5.0
7
5.0
8
5.0
9
5.0
10
5.0
34 / 120
Maximum Time
Nominal Occlusion
to occlusion alarm
Alarm Pressure
(min:sec)
(mmHg)
2:00
8:00
20:00
184
33:00
276
52:00
368
65:00
460
85:00
552
102:00
664
120:00
736
140:00
828
155:00
920
01:00
02:00
05:00
184
07:00
276
10:00
368
12:00
460
15:00
552
17:00
644
20:00
736
24:00
828
26:00
920
Maximum Infusion
Pressure
(mmHg)
0
50
92
110
220
330
450
560
670
780
80
1000
1100
0
50
92
110
220
330
450
560
670
780
890
1000
1100
Maximum Bolus
Volume
(ml)
0.1
0.2
0.3
0.5
0.7
0.9
1.0
1.2
1.6
1.8
2.0
0.1
0.2
0.3
0.5
0.7
0.9
1.0
1.2
1.6
1.8
2.0
5001FAOPT71 ISS 8.0