INDICATOR LED TEST: Check that the STOP and START LED' s are
flashing.
DECLUTCH SWITCH TEST: Squeeze plunger holder finger grips and
check that the display alternates between 1 (engaged) and 0
(disengaged - finger grips squeezed together).
NEAR END OF INFUSION TEST: Starting with an empty, extended
50/60ml syringe on the unit, squeeze the finger grips and move the
plunger holder slowly to the left.
Check the display switches from 0 to 1. The position at which the display
changes will depend on the syringe type being used. ( For example 6ml
on a B-D Plastipak 50/60ml syringe and 3ml on a Terumo 50ml syringe -
see Technical Service Manual for values).
LINEAR GRID TEST: Squeeze the finger levers and position the drive
mechanism to the right then slowly move the syringe plunger to the left
observing the display alternate between 0 (OPTO over slot) and 1 (OPTO
over bar).
PLUNGER DETECTOR TEST: Press plunger plate button. Check the
display switches from 1 ( No syringe fitted) to 0 ( Syringe plunger fitted).
MOTOR ENCODER TEST: Motor is pulsed while encoders are tested.
COVER DETECT TEST: Open and close the cover. Check that display
changes from 0 (cover open) to 1 (cover closed).
SYRINGE SIZE DETECTION SYSTEM: Lift syringe clamp and check
that the number shown in the display increases within the normal range
(approx. 045 to 215).
PUMPING PRESSURE DETECTION TEST:
confirm that value is within normal range (-020 to +020). Gently press
back on the plunger holder and observe the value increase.
PATIENT HAND SET TEST: Connect patient hand set. Check display
switches from OFF to ON as the button is pressed.
KEY SWITCH TEST: Turn key from SET position to RUN position. Check
display changes.
Summary of user defined options selected.
Returns unit to normal operation.