For technician access codes please refer to the Technical Service Manual.
5001FAOPT71 ISS 8.0

Configured Options

Drug Names and Safety Limits

1. Turn key switch to SET position while
pressing down START button.
2. Enter the access code using the rate
selection arrows.
3. Select
DRUG NAMES AND SAFETY
LIMITS from the menu. Press ENTER.
4. Press MODIFY DRUG to modify the drug
summary. Use "+" and "-" arrows to select
desired values. When field is correct press
OK to store the selection.
5. NAME: Use "+" and "-" arrows to set
highlighted letter. Press NEXT for next
character (up to ten letters). Press OK
when complete.
6. MIN DRUG CONC: Use "+" and "-" arrows
to set minimum concentration. Press OK
when complete.
7. MAX DRUG CONC: Use " +" and "-" arrows
to set maximum concentration. Press OK
when complete.
8. MIN LOCKOUT PERIOD: Use "+" and "-"
arrows to set minimum lockout period.
Press OK when complete.
9. MAX LOCKOUT PERIOD: Use " +" and " -"
arrows to set maximum lockout period.
Press OK when complete.
10. MIN PCA DOSE: Use " +" and "-" arrows to
set minimum PCA dose. Press OK when
complete.
11. MAX PCA DOSE: Use "+" and "-" arrows to
set maximum PCA dose. Press OK when
complete.
12. MAX CONTINUOUS: Use "+" and "-" to set
maximum continuous rate. Press OK when
complete.
17 / 120