Getting Started

PCAM Patient History

Each time the PCAM unit is switched ON it will ask if this is a new patient. Pressing YES will
provide opportunity to re-set patient history. Pressing NO continues with the current protocol and
retains all protocol records, event history, graphs etc.
The PCAM will retain the events in a rolling memory. Following selection of a new patient, it
remains possible in technician mode to access previous patient(s) history still held in memory.
Patient history can be accessed at any time by pressing the HISTORY button. PCAM provides a
clear rolling 24 hour graphical representation of the PCA demand pattern and the drug
administered to the patient. The graphs are updated when the history button is pressed and give
values for each completed hour and the current hour.

PCA Demands

Provides a record of the last 24 hours good and failed PCA demands. The good demands are
indicated by the shaded section of the graph and the failed demands by the clear section. The
latest hour is shown at the right side of the display.
This graph provides a clear picture of good and bad PCA demands and pattern of the patients
usage. Used in conjunction with the PCA Demands graph, this display helps to indicate if the PCA
protocol needs modification and when to end treatment.
1. To access the PCA demand graph press
the HISTORY button once.
2. To exit the screen press the BACK softkey.
3. To scroll through to the next History screen
press the HISTORY button.

Drug Infused

Record of the total amount of drug administered to the patient over the last 24 hours. The latest
hour is shown at the right side of the display. This graph provides a clear picture of the actual
drug administered, including loading dose, continuous background infusions, clinician over-rides
and protocol changes.
Used in conjunction with the PCA Demands graph, this display helps to indicate relative pattern
of the demand pattern and the actual drug administered.
1. To access the PCA demand graph press
the HISTORY button twice.
2. To return to the previous screen press the
BACK softkey.
3. To scroll through to the next History screen
press the HISTORY button.