Getting Started
Operating Precautions
This instrument is protected against the effects
of external interference, including high energy
radio frequency emissions, magnetic fields
and electrostatic discharge (for example, as
generated by electrosurgical and cauterising
equipment, large motors, portable radios,
cellular telephones etc.) and is designed to fail
safe if unreasonable levels of interference are
encountered.
In some circumstances the unit may be
affected by an electrostatic discharge of +/-8kV
(contact), +/-15kV (air), at test levels lower
than these values the unit will operate
normally. In rare circumstances the unit may
be affected by radiation at a level of 10V/m. If
the unit is affected by this external interference
the unit will fail safe or reset, ( a call back alarm
will occur after 2 minutes). Should false alarm
conditions be encountered either, remove the
source of the interference, or regulate the
infusion by another appropriate means.
If this instrument is dropped, subjected to
excessive
moisture,
humidity
or
high
temperature, or otherwise suspected to have
been damaged, remove it from service for
inspection by a qualified service engineer.
An explosion hazard exists if the instrument is
used
in
the
presence
of
flammable
anaesthetics. Exercise care to locate the unit
away from any such hazardous sources. An
electrical shock hazard exists if the units
casing is opened or removed. Refer all
servicing to qualified service personnel.
A comprehensive service manual containing
circuit descriptions, servicing and testing
information is available for this unit. It can be
ordered from your ALARIS Medical Systems
authorised
distributor
(Technical
Service
Manual Part Number 5000PB00004).