Operation of PCAM is greatly simplified by the use of PRE-SET PCA protocols. When the PCAM
key switch is turned to the SET position the unit will automatically display pre-set PROTOCOL A
if NEW PATIENT has been selected, or, display the previous protocol in use if NEW PATIENT has
not been selected.
With the key switch in the SET position, it is possible for the user to modify the pre-set protocol
using the MODIFY PROTOCOL button and select another pre-set protocol using the NEXT
PROTOCOL button.
To modify a Pre-Set Protocol
1. Press MODIFY PROTOCOL indicated on
the screen.
2. The protocol summary will be displayed.
Use ARROWS to highlight a field, press
ALTER to enter that field and "+" , " -"
arrows to select desired values.
3. When field is correct press CONFIRM or
CANCEL.
4. Display will show:
5. Press OK to return to display.
(Note:
A modified protocol has no pre-set
letter).
To change to another pre-set protocol
Press NEXT PROTOCOL to display the step through the pre-set protocol.
Pre-set protocols are identified as A to E. When the desired protocol has been selected it can be
used by turning the key switch to the RUN position, or, can be modified using the MODIFY
PROTOCOL.
To confirm protocol
When ever a new protocol is selected, modified or the key switch is turned to the SET position,
the CONFIRM PROTOCOL display will appear. Pressing OK automatically records the protocol
and any changes in the EVENT LOG history.
14 / 120
Drug 'X'
Protocol
A
Drug name
Drug conc.
PCA Dose
Lockout period
Occlusion level
Continuous
Loading dose
Max. limit
Limit duration
PCA delivery
Drug ' X '
1.0mg/ml
1.0mg
(1.0ml)
5 mins
4
0µg/h
(0.0ml/h)
0µg
(0.0ml)
50mg
(50.0ml)
4 hours
STAT
5001FAOPT71 ISS 8.0