Specifications
OPERATION MODE:
Continuous
CRITICAL VOLUME:
The maximum over infusion which can occur in the event of a single fault condition is 0.8ml for
20ml, 30ml and 50ml syringes and 1.5ml for a 100ml syringe.
ALARM CONDITIONS:
Pressure Limit exceeded
Drive Disengaged
Syringe Almost Empty
Syringe Empty
Internal Malfunction
Nurse Attention/Call Back
PUMPING PRESSURE / ALARM LEVEL:
375mmHg (nominal) default alarm level (L-4) with 11 user selectable alarm levels (L-0 to L-10).
Syringes may limit below level 10.
(The maximum pressure that can be developed by the system at the maximum user selectable
alarm level is 1100mmHg).
CLINICIAN OVER-RIDE:
Bolus or continuous infusion in RUN mode.
(User selectable from 1mg - 99.9mg or 0.1ml to 99.9ml (volume mode) bolus dose administered
at the STAT rate (100ml/h) or over 1 to 180 minutes delivery period).
Modify PCA Protocol in SET mode.
(When option to disable MODIFY PROTOCOL has been selected).
BATTERY OPERATION:
6 hours operation from a fully charged battery at 5.0ml/h and 20°C under normal conditions.
BATTERY TYPE AND RECHARGE TIME:
Rechargeable sealed lead acid type. 10 hours from discharge to 80% charge, 24 hours from
discharge to 100% charge.
EVENT HISTORY:
1500 events rolling memory.
MEMORY RETENTION:
All calibration and set up information will be retained in the pump memory for a minimum of 3
years.
30 / 120
Low Battery Warning
Battery Exhausted
Cover Open during operation
Hand-set Disconnected
Syringe Error
5001FAOPT71 ISS 8.0