IVAC P5000 Mk II Mode D'emploi page 24

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Configured Options
Drug Names and Safety Limits
13. MAX LOADING DOSE: Use "+" and " -"
arrows to set maximum loading dose.
Press OK when complete.
14. MAX DOSE LIMIT: Use "+" and "-" arrows
to set maximum dose limit. Press OK when
complete.
15. MAX CLINICIAN BOLUS: Use "+" and "-"
arrows to set the maximum clinician bolus.
Press OK when complete.
16. NEXT DRUG to display the next drug
name and the safety limits. The five pre-set
drug protocols are identified as 1 to 5.
17. Press QUIT
to
configuration menu.
18. When set-up is complete, turn key switch
to OFF position to save selection and to
turn PCAM off.
18 / 120
exit
and
return
to
5001FAOPT71 ISS 8.0

Publicité

Table des Matières
loading

Produits Connexes pour IVAC P5000 Mk II

Table des Matières