IVAC P5000 Mk II Mode D'emploi page 22

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Getting Started
Alarm Procedures
Description
Display
BATTERY CHARGE LOW WARNING
Battery charge low with 30 minutes operation
remaining. Battery indicator will flash and after 30
BATTERY LOW
minutes a continuous audible alarm will indicate that the
battery is exhausted. Reconnect to AC power supply to
continue operation and charge internal battery.
BATTERY EXHAUSTED
Internal battery exhausted. To silence the alarm switch
the key switch to the OFF position and reconnect unit to
BATTERY EXHAUSTED
AC power supply. Restart operation on AC power whilst
charging the internal battery. Switch to the RUN
position.
NEAR END OF SYRINGE WARNING
Syringe almost empty with about 6% of its volume
SYRINGE NEAR EMPTY
remaining. Press START to silence alarm and continue
operation. Display will flash SYRINGE NEAR EMPTY.
SYRINGE EMPTY - END OF INFUSION
The pump has reached the end of the infusion. About
SYRINGE EMPTY
1% of the syringe volume will remain in the syringe
helping to prevent the infusion of air bubbles into the
PCA set.
AC POWER SUPPLY DISCONNECTED WARNING
AC Power has been disconnected and the pump is
operating on battery power. Reconnect AC power
AC POWER FAIL
supply or press STOP to silence the alarm and continue
battery operation. The display will light up ON
BATTERY . The alarm will automatically cancel if the AC
power supply is reconnected.
INTERNAL MALFUNCTION
The alarm system has detected an internal malfunction.
MALFUNCTION
Note the malfunction code. Remove unit from service
for examination by a qualified service engineer.
Alarm Procedures - Alarms without screen prompts
Description
NURSE ATTENTION WARNING
3 BEEPS
Unit left switched ON for over 2 minutes without
starting operation. Press STOP or any of the control
buttons to silence the alarm for a further 2 minutes.
16 / 120
5001FAOPT71 ISS 8.0