Alarms are indicated by a combination of an audible alarm, flashing amber STOP light and a
descriptive message in the display.
Display
COVER OPENED
DRIVE DISENGAGED
LINE OCCLUSION
SYRINGE ERROR
CHECK HANDSET
5001FAOPT71 ISS 8.0
1. A continuous audible alarm indicates that
the infusion has stopped. First press MUTE
to silence the alarm for a maximum of 2
minutes, then check the display for an
alarm message. Press STOP to cancel the
alarm message.
2. When the cause for the alarm has been
rectified press START to restart operation.
Description
COVER OPENED DURING OPERATION
The cover has been opened, or cover lock operated,
during operation. Check cover and lock.
PUMP DRIVE DISENGAGED
The drive system has been disengaged during
operation. Unlock and open the cover. Check the finger
grips and the position of the syringe.
EXCESSIVE PUMPING PRESSURE
Pumping pressure has reached the alarm limit. Unlock
and open the cover, squeeze finger grips on the plunger
holder to release the drive mechanism and relieve any
excessive pressure in the syringe and patient line.
Identify and remove the cause of the blockage in the
drive,
syringe,
or
restarting the infusion.
SYRINGE SIZE ERROR / FITTED INCORRECTLY
Incorrect size of syringe has been fitted, the syringe
clamp has not been positioned correctly on the syringe
or has been disturbed during operation or plunger is not
fitted in plunger slot. Unlock and open the cover, check
syringe size, position of syringe clamp, syringe and
plunger.
PATIENT HAND SET FAILURE
Patient hand set has become faulty or disconnected
during operation. Check operation and connection of
the hand set to the unit. Press START to continue if
operation without the hand set is required.