Front Panel And Main Display; Purge; Clinician Over-Ride - IVAC P5000 Mk II Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Pump Status
Demand
Status
Current Protocol & Syringe Type
The PURGE button allows the delivery of a limited volume of fluid in order to purge the extension line prior to
being connected to a patient.
The purge feature is not activated when the cover is closed. Ensure that the extension line
is disconnected from the patient before purging the line. No alarms are disabled during the
operation of the purge feature.
The clinician over-ride feature can be used in RUN mode to administer an additional bolus dose or continuous
background infusion of a limited dose and duration, for example during the PCA lock out period. The clinician
over-ride is a special feature which can be configured according to the specific clinical situation. Clinician over-
ride can also be used in SET mode to allow modification of the pre-set PCA Protocol when this option has been
disabled for normal use.
The delivery of the clinician over-ride continuous infusion will automatically halt while a
Patient or Clinician over-ride bolus is being administrated. To cancel clinician over-ride
during delivery, press STOP and press the YES softkey.
5001FAOPT71 ISS 8.0

Front Panel and Main Display

Drug Infused
1. PURGE can only be operated with the cover
opened and the key switch in the RUN position.
2. Press the PURGE buttons together until fluid
flows and priming of the syringe extension line is
complete. The audible alarm will operate during
use of the PURGE buttons and the volume used
during priming will be shown in the volume
infused display.
:
Important
1. Turn key to RUN position and ensure green light is
illuminated on the START button.
2. Press and hold down the CLINICIAN OVER-RIDE
button for 2 seconds.
3. Use "+" and "-" arrows and NEXT button to enter
three figure pre-programmed clinician access code
"n n n". See technical service manual.
4. Select BOLUS or CONTINUOUS.
5. Use "+" and "-" arrows to select the dose delivered,
when the correct value has been entered press
OK.
6. Use "+" and "-" arrows to select period over which
the dose is to be delivered. Press OK when correct
time has been entered.
7. BEGIN BOLUS?
continuous infusion will be delivered to the patient.
NO - Quit set up and return to normal operation.
Important:
Getting Started
Time
Pumping
Pressure
Maximum
Dose

Purge

Clinician Over-ride

YES
-
Clinician
bolus
9 / 120
/