The self test routine is designed to allow confirmation of many of the pump functions, defaults and
calibrations without requiring internal inspection. It does not represent a full calibration check.
1. Turn key switch from OFF to SET while
holding down the START button.
2. Enter "1 2 3" using the "+" , "-" and NEXT
buttons.
3. Press ENTER
to commence self test
sequence.
4. The unit will now proceed through a series
of tests as described below.
5. Press NEXT button to move to next test.
Should the pump fail the test sequence, it should be taken out of service and
inspected by a qualified service engineer.

Self Test Sequence

Display
1.
SOFTWARE
REVIEW
2.
SYRINGE REVIEW
3.
-:- - v
4.
" BEEP"
5.
6.
Press S1
24 / 120
Important:
Test Description
SOFTWARE/SET UP REVIEW: Including software version, program crc,
language, communications fitted.
REVIEW SET UP OF SYRINGE CALIBRATION DATA: Including syringe
calibration, occlusion calibration and syringe range.
INTERNAL PSU VOLTAGE:
AUDIBLE ALARM TEST: Check alarm sound continuously.
DISPLAY AND BACKLIGHT DIM TEST: Check that all the display is
operating and that the display operates on a full backlight for the first part
of the test, dims for the second and then switches off at the end of the
test.
TOUCH PANEL SWITCH TEST: Press buttons in turn from S1 to S10
form START to PURGE until end of test is reached.
5001FAOPT71 ISS 8.0