Troubleshooting

10
Troubleshooting
Please review the checklist below should you experience a problem
with your Propex Pixi
proposed solutions, please contact your distributor.
Warning
The following patient related factors may prevent accurate readings:
- Blocked root canals;
- Teeth with large apices;
- Root fracture or perforation;
- Metal crowns or bridges, if they come into contact with the file or
the lip clip.
#
Problem
During battery charging
the battery symbol blinks
at a fast rate.
1
32/222
®
. If the problem persists after following the
Possible cause
The battery is not
connected.
The battery is not
rechargeable.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Solution
Open the battery
compartment, and connect
the battery as described in
the User Manual, section
6.4.
Replace the battery with a
rechargeable type as
described in the user
manual, section 6.4.