TROUBLESHOOTING

NO TABLE ROTATION OR TABLE IS NOT IN THE DOWN POSITION AT THE START OF THE PROCESS.
ERROR 1
Indicates table rotation was not detected, or table is not in the down position at start of menu.
Make sure the table handle is pushed completely forward and the lift is in the DOWN position, then push
ESC. If the lift is still elevated, turn OFF the unit, open the door and gently push the table down to the
stop position. Turn the unit ON.
OVER TEMPERATURE
ERROR 2
The temperature of the unit has exceeded the maximum temperature allowed.
Turn unit OFF then ON.
Check blower operation.
Check fan operation.
BULB FAILURE
ERROR 3
One or more bulbs failed.
The bulb layout seen on the display is the same as arranged in the unit, as viewed from the top.
The failed bulb will blink on the display.
Follow the bulb replacement procedure in the Maintenance Section.
NOTE: A power surge may cause any of the above errors to appear. Follow the
instructions provided to clear the error and resume normal operations.
If any error persists, call Dentsply Sirona service department for assistance.