BULB AND FILTER REPLACEMENT (CONTINUED)
1.
Turn the power switch to the OFF position.
3. Loosen the two screws on rear of top panel.
5. Put supplied gloves on hands.
!
Failure to wear gloves could cause premature lamp failure due to skin oils.
7.
Place center bulb bracket in slots on left side of internal cover. Be careful not to remove the filters in
the bracket. Do not touch the filters with bare hands, handle with gloved hands only. If filters are being
replaced, the filters have a frosted side and a shiny side. The frosted side should be placed toward the
lamp. ONLY USE REPLACEMENT FILTERS SUPPLIED BY DENTSPLY SIRONA.
2. Unplug the AC power cord from
the wall outlet.
4 Slide top panel backwards until full
access to bulbs is achieved.
6. Loosen center bulb bracket retaining
screws, if needed, for bulb clearance.
Loosen 2 screws.
Frosted Side UP!
11
Front of unit