6 Troubleshooting

6
Troubleshooting
Fault
LED
flashing
Door cannot be closed properly
LED is off
Handpiece maintenance cycle
does not start
Top of oil can is more than half
filled with oil
No oil at handpiece chuck after
maintenance cycle
Possible cause
Spray can failure
Compressed air not connected
or not on/flowing
Handpieces placed incorrectly
No power supply
Front door not closed
O-Ring worn or damaged
Clogged handpiece
Oil system malfunction
11
Remedy
Re-insert or replace spray can
Turn on / connect compressed air
Handpiece chucks must be directed towards
the front or back
Plug in power cord
Turn unit on
Close front door
Contact service (chapter 11)
Consult handpiece instructions
Contact service (chapter 11)
®
M
A
®
IDWEST
UTOMATE
Automated Handpiece Maintenance System
EN