Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Immunity test
Voltage dips
Voltage
interruptions
IEC 61000-4-
11
Rated power
frequency
magnetic fields
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
IEC 60601-1-2
Test level
0% UT;
0% UT;
0.5 cycle
0.5 cycle
0% UT; 1 cycle
0% UT; 1 cycle
and
and
70% UT;
70% UT;
25/30 cycles
25/30 cycles
0% UT;
0% UT;
250/300 cycles
250/300 cycles
30 A/m
30 A/m
50 or 60 Hz
50 or 60 Hz
Note
UT is the a.c. mains voltage prior to application of the test
level.
Compliance
level
Mains power quality should be
that of a typical public low-
voltage power supply network
that supplies buildings used
for domestic purposes,
commercial or hospital, clinical
environment. If the user of the
Product requires battery
charging during power mains
interruptions, it is
recommended that the
Product charger be powered
from a separate power supply
(UPS, etc.).
Power frequency magnetic
fields should be at levels
characteristic of a typical
public low-voltage power
supply network that supplies
buildings used for domestic
purposes, commercial or
hospital, clinic environment.
Electromagnetic
environment - guidance