Cleaning, Disinfection and Sterilization
Operation
5
Disinfection
6
Rinsing
28/222
Operating mode
Once the accessories
have been cleaned
and inspected, place
them into the
disinfection bath for
the prescribed
contact time (e.g.
Cidex OPA, Johnson
& Johnson Medical,
100% for 20 minutes);
the accessories must
be sufficiently
immersed in the
solution.
Remove the
accessories from the
disinfection bath and
rinse them off
thoroughly at least
five times for a minute
each time with water.
Dry the accessories
by blowing them
down fully.
Abundant rinsing
(at least 1 min).
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Warning
The hook mechanism has to
be activated several times
during disinfection and the
rinse to allow the inner parts
to be disinfected more
effectively.
To dry use oil-free, filtered
compressed air and then
leave the accessories to dry
further in a clean place for at
least 20 minutes.
Once the accessories are
dry, inspect and pack them
as soon as possible.
Use quality water in
accordance with local
regulations.
If a disinfecting solution
contains a corrosion
inhibitor, it is recommended
to rinse the instruments just
before the autoclaving.
Dry on a single use non-
weaved cloth, or with a
drying machine or filtered
compressed air.