Technical Characteristics; Error Code - Dentsply Sirona Propex Pixi Manuel De L'utilisateur

Masquer les pouces Voir aussi pour Propex Pixi:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38

Technical Characteristics

8
Technical Characteristics
Propex Pixi
EMC (Electromagnetic compatibility) standards.
Propex Pixi
of medical devices:
Internally powered equipment (AAA 1.2V 1000mAh NiMH
rechargeable battery);
Type BF applied parts;
Not suitable for use in the presence of flammable anesthetic
mixtures with air, oxygen or nitrous oxide;
Continuous operation;
Ingress of liquids – not protected;
Environmental conditions during storage/transportation:
- Temperature: –20°C to +60°C (0°F to 140°F);
- Relative humidity: 10% to 90%, non-condensing;
- Atmospheric pressure: 106 kPa to 19 kPa.
Environmental conditions during device usage:
- Temperature: 10°C to +40°C (50°F to 104°F);
- Relative humidity: 10% to 90%, non-condensing;
- Atmospheric pressure: 106 kPa to 70 kPa.
Specifications:
Dimensions
Weight
Type of screen
Supply
Switching charger
9

Error Code

None.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
®
complies with IEC 60601-1 Safety and IEC 60601-1-2
®
electronic apex locator belongs to the following category
66 x 55 x 18 mm
55 gr.
Color LEDs
AAA 1.2V 1000mAh NiMH rechargeable battery
Input: 100-240 V AC ~ 50-60 Hz
Output: 6V DC ± 5%, 1000 mA
31/222

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

A1030 000 001 00

Table des Matières