Apex Localization; Getting Started - Dentsply Sirona Propex Pixi Manuel De L'utilisateur

Masquer les pouces Voir aussi pour Propex Pixi:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
Step-by-Step Instructions
6.6

Apex Localization

6.6.1

Getting Started

Disconnect the charger from the device if connected.
1. Before connecting the measurement cable with attached lip clip and
connection hook to the patient, plug measurement cable into the
device and turn on the device by pressing the
on the top of the device. The first bar will start blinking.
2. Attach the lip clip to the patient.
3. Gently insert the file into the canal.
Note
To ensure optimal performance the file size should be adjusted to the
canal diameter.
4. Connect the connection hook to the metal shaft of the file.
The first bar will stop blinking accompanied by a double beep signal –
see
Fig.
Note
A blinking bar indicates faulty connection. Disconnect the
measurement cable from the patient and check cable connections,
clean the connection hook and the lip clip, moisten the canal if
necessary and start again.
No other adjustments are required before starting apex localization.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
8.
Fig. 8
"ON / OFF" button
17/222

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

A1030 000 001 00

Table des Matières