ResMed VPAP III ST-A Manuel Utilisateur page 33

Masquer les pouces Voir aussi pour VPAP III ST-A:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 117
Function
Default
SmartStart
OFF
Mask Alarm
ON
SmartStart™
The VPAP III ST-A has a function called SmartStart which can be enabled by your
clinician. If SmartStart is enabled, VPAP III ST-A will start automatically when you
breathe into the mask and will stop automatically when you take your mask off.
This means you do not have to press the Start/Stop key to begin or end
treatment.
Note: If you select "Mir Full" as the mask option, SmartStop is automatically
disabled. SmartStart may not work with a full face mask due to safety features
of the mask.
When Mask Alarm is set to ON, SmartStart/Stop automatically reverts to OFF.
SmartStop cannot be used with Mask Alarm because if a high leak occurs,
SmartStop will stop treatment before the Mask Alarm signal is activated.
Settings for Mask Types
The following table shows the setting that should be selected for each mask
type.
Table 3: Settings for mask types
Settings
MIRAGE
ULTRA
STANDARD
MIR FULL
Function Description
Turns the SmartStart function on
or off.
See "SmartStart™" on page 27 .
Turns the Mask Alarm signal on or
off.
If enabled, the Mask Alarm feature
will alert you when a high mask
leak is detected. An audible tone
will sound and a high leak
message will appear on the LCD
screen.
Mask
Mirage Swift Nasal Pillows System
Mirage Nasal Mask
Ultra Mirage Nasal Mask
Mirage Activa Nasal Mask
Mirage Vista Nasal Mask
Modular Nasal Mask
Mirage Full Face Mask
Mirage Full Face Mask Series 2
Ultra Mirage Full Face Mask
Setting Options
ON/OFF
ON/OFF
How to Use the VPAP III ST-A Menus
27

Publicité

Table des Matières
loading

Table des Matières