Warnings; Cautions - ResMed VPAP III ST-A Manuel Utilisateur

Masquer les pouces Voir aussi pour VPAP III ST-A:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 117
hypotension or significant intravascular volume depletion
pneumothorax or pneumomediastinum
recent cranial trauma or surgery.

Warnings

The entire manual should be read before using the VPAP III ST-A.
Advice contained in this manual should not supersede instructions given by
the prescribing physician.
The VPAP III ST-A should be used with masks and accessories recommended
by ResMed or the prescribing physician. Use of incorrect masks and
accessories may adversely affect the function of the VPAP III ST-A.
The VPAP III ST-A is designed for use with masks that allow exhaled gases to
be flushed out through vent holes. Exhaled gases will be rebreathed if the
mask is worn with the machine turned off, or the vent holes are occluded. If
this occurs over prolonged periods, suffocation may occur.
In the event of power failure or machine malfunction, remove the mask.
The VPAP III ST-A can be set to deliver pressures up to 30 cm H
unlikely event of certain fault conditions, pressures up to 40 cm H
possible.
The VPAP III ST-A is not suitable for use in the vicinity of flammable
anesthetics.
The VPAP III ST-A should not be used with anesthetized patients, whose
breathing depends entirely on mechanical ventilation.
If oxygen is used with the VPAP III ST-A, the oxygen flow should be stopped
when the device is not operating. If oxygen flow continues when the device
is not operating, oxygen may accumulate within the device and create a risk
of fire.
Do not use the VPAP III ST-A if there are obvious external defects,
unexplained changes in performance or unusual noises.
Do not open the VPAP III ST-A case. There are no user serviceable parts
inside. Repairs and internal servicing should only be performed by an
authorized service agent.

Cautions

At low EPAP pressures, the flow through the mask vent holes may be
inadequate to clear all exhaled gases, and some rebreathing may occur.
The air flow for breathing produced by this device can be as much as 6
higher than the temperature of the room. Caution should be exercised if the
room temperature is warmer than 32
Note: The above are general warnings and cautions. Further specific warnings,
cautions and notes appear next to the relevant instructions in the manual.
6
o
C.
O. In the
2
O are
2
o
C

Publicité

Table des Matières
loading

Table des Matières