Continuous Operation; Display Resolution - ResMed VPAP III ST-A Manuel Utilisateur

Masquer les pouces Voir aussi pour VPAP III ST-A:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 117
IEC 60601-1 Classifications
Class II (double insulation)
Type CF

Continuous operation

This flow generator is not suitable for use in the presence of a flammable
anesthetic mixture with air, or with oxygen or nitrous oxide entrained in the flow
generator airpath.
Table 7: Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
** The displayed values are estimates. They are provided for trending purposes only.
42
Range
-5 to 30 cm H
0–120 L/min
100–3000 mL
6–60 BPM
0.6–60 L/min
Results may be inaccurate in the presence of leaks.
Accuracy
O
±0.5 cm H
O
2
2
**
**
±0.5 BPM
**

Display Resolution

0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min

Publicité

Table des Matières
loading

Table des Matières