Operator Control Panel

4.0 Operator Control Panel
See figure 2, page 5.
MACHINE STATUS
FLOW-OK indicates water is flowing
adequately.
WARMING indicates the machine is
heating.
COOLING indicates the machine is
cooling.
TEST LIGHTS flashes all indicator lights
and digital displays, while sounding an
audible alarm.
ALARM SILENCE temporarily silences
the audible alarm. The audible alarm will
reactivate in 10 minutes if the condition
persists.
4
ALERTS
CHECK FLOW indicates the flow of
water to the blanket/body wrap has been
restricted.
NOTE: When two gray hose sets are in use,
the CHECK FLOW alarm will only sound if
the flow through BOTH gray hose sets is
occluded. (See page 7, section 5.6).
CHECK PATIENT indicates the patient
probe is sensing an abnormal patient
temperature (below 29°C or above 40°C).
It may indicate the patient probe has
become dislodged, an incorrect probe
is being used, or a probe has become
defective. An audible alarm accompanies this
alert.
ADD WATER indicates the water
reservoir is filled to less than the
recommended level.
REMOVE FROM USE NOW indicates
the Medi-Therm III machine has shut down
due to a malfunction.
Remove the machine from use.
Contact qualified medical service
personnel.
FLASHING DISPLAY indicates that the
water or patient temperature is not being
maintained within the acceptable limits of
the set point temperature. It is accompanied
by an audible alarm. See pp. 8-10, sections
6.1 and 6.2 for explainations of the various
alarms that have a flashing temperature
display.
POWER LOSS INDICATOR
lights when power is removed and
ON/OFF circuit breaker is ON.
[located on front of machine]