Care & Cleaning
10.0 Care and Cleaning
As with any precision equipment, proper care and
maintenance will insure proper performance and
reliability. The following routine maintenance procedures
will help insure dependable Medi-Therm III operation.
10.1 Machine Care
CAUTION
• Fill with distilled water only.
Failure to use distilled water may result in
poor machine performance.
• Do not use alcohol. Alcohol may accelerate
blanket/body wrap deterioration. Alcohol is
flammable.
• Do not operate without water, since damage to
internal components may result.
• Do not overfill. Overfilling may result in
overflow because the water in the blanket/body
wrap drains back into the machine when the
machine is turned off.
• When filling an empty Medi-Therm III machine,
add a small amount of germicidal or sanitizer
solution to the water (Air-Kem A-33 liquid or
equivalent). Carefully follow the manufacturer's
recommended concentration to prevent algae
growth within the Medi-Therm III system and
Hyper/Hypothermia blanket/body wrap.
Do not use too much concentrate, since it
will result in "foaming," which can reduce the
blanket/body wrap efficiency and may damage the
circulating pump.
• Do not use bleach. Bleach will damage the
heating element.
10.2 Draining the Machine
It is not necessary to drain the Medi-Therm III
machine after each use. However, periodic draining is
recommended at least once a month.
14
10.3 Machine Cleaning
CAUTION
Users should not use cleaning or decontamination
methods different from those recommended.
To clean the external surfaces of the machine and
connector hose, use a nonabrasive cleaning solution
(such as warm, soapy water) and a clean cloth.
Wipe or air dry. Apply a disinfectant such as 10%
chlorinated bleach solution (chlorinated bleach with
5.25% sodium hypochlorite) to the external surfaces
and allow to dry.
10.4 Blanket/Body Wrap Care
CAUTION
• Do not store filled blankets/body wraps. Algae
growth may occur inside.
• Do not use a blanket/body wrap if severe algae
buildup occurs.
• Do not expose blankets/body wraps to
temperatures over 65°C.
Disposable Blankets/Body Wraps
1.
After the machine has been turned off, raise the
blanket/body wrap higher than the machine to
allow most of the water to drain back into the
machine.
2.
Close both sets of hose clamps.
3.
Disconnect the blanket/body wrap from the
machine.
4.
Dispose of blanket/body wrap.
Reusable Blankets
1.
After the machine has been turned off, raise the
blanket higher than the machine to allow most of
the water to drain back into the machine.
2.
Close both sets of hose clamps.
3.
Disconnect the blanket from the machine.
4.
Manually clean both sides of the blanket on a
flat surface with warm water, a mild commercial
detergent, and a sponge or cloth.
Medi-Therm
III
®