Operator Control Panel; Machine Status - Gaymar Medi-Therm II Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20

Operator Control Panel

4.0 Operator Control Panel
See figure 2, page 5.

MACHINE STATUS

HEAT indicates the machine is heating.
IN-TEMP indicates that the blanket or
patient temperature is at the desired
temperature setting.
COOL indicates the machine is cooling.
TEST LIGHTS flashes all indicator lights
and digital displays, while sounding an
audible alarm.
ALARM SILENCE temporarily silences
the CHECK FLOW, CHECK PATIENT PROBE,
FLASHING DISPLAY, and ADD WATER alarms
for 10 minutes.
4
ALERTS
CHECK FLOW indicates the flow of
water to the blanket has been restricted.
CHECK PATIENT indicates the patient
probe is sensing a temperature below 29°C
or above 40°C.
ADD WATER indicates the water
reservoir is filled to less than the
recommended level.
REMOVE FROM USE NOW indicates
the Medi-Therm II machine has shut down.
Remove the machine from use.
FLASHING DISPLAY indicates that
the blanket or patient temperature is not
being maintained within 0.5°C of the set
point temperature. It is accompanied by an
audible alarm.
POWER LOSS INDICATOR
lights when power is removed and
ON/OFF circuit breaker is ON.
[located on front of machine]
Medi-Therm
TM
II

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mta6012ce

Table des Matières