Medi-Therm
TM
6.
Air dry or wipe with a clean cloth.
7.
Apply a disinfectant such as a 10% chlorinated bleach
solution (chlorinated bleach with 5.25% sodium
hypochlorite) to both sides of the blanket.
8.
Allow to air dry. Solution contact time is what makes
disinfection effective.
NOTE: Excess solution pooled in the buttons can
be removed with a clean, dry cloth.
9.
Loosely roll up the blanket and store for next use.
Reusable blankets can be cleaned and reused so
long as they do not crack or leak.
10.5 Probe Care
Reusable Probes
For cleaning, disinfecting, and sterilizing reusable Gaymar
probes (or equivalent YSI 400 series reusable probes),
refer to the instructions provided with the probe.
Disposable Probes
Gaymar disposable probes should be discarded after use.
10.6 Probe Check Well Care
pr obe
probe
check
check
w el l
well
pr ot ect i ve
protective
sheat h
sheath
Figure 8—Probe Check Well
The PROBE CHECK WELL should be cleaned with a
small tubular brush as outlined in section 10.3, Machine
Cleaning.
II
pr obe
probe
Care and Cleaning

10.7 Quick-Disconnects

The quick-disconnects joining the connector hose to the
machine may become difficult to engage. This problem
can be prevented by periodically applying a silicone base
or light machine oil to the outside of the male connector
prior to engagement.
10.8 Preventive Maintenance and Service
Preventive maintenance and service procedures are
described in the Medi-Therm II MTA6012CE Service
Manual. These procedures should be performed annually
(or as required by hospital protocol) by qualified
personnel such as certified biomedical electronics
technicians or certified clinical engineers.