Experiment Set-Up - 3B SCIENTIFIC PHYSICS U104001 Instructions D'utilisation

Appareil d'analyse du point critique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
and let in SF
till a pressure of approx. 10 bar has
6
been attained.
Shut the regulating valve.
Open the flush valve slightly till the pressure has
dropped to almost 0 bar.
Shut the flush valve.
Filling with test gas:
After at least four flush cycles, press the MINI-
CAN® gas container against the gas connection fit-
tings. Slowly open the regulating valve and let in
SF
till a pressure of approx. 10 bar has been at-
6
tained.
Shut the regulating valve.
Wind the handwheel in the opposite direction till
the piston reaches a position of say 46 mm.
Press the MINICAN® gas container against the gas
connection fittings, slowly open the regulating
valve and shut it again when a pressure of 10 bar
has been attained.
7.4 Recommendation for storage lasting for short
periods of time:
One gas filling can remain in the measuring cell for
several days.
If no experiments are being conducted, wind the
handwheel back till the piston is in a position where
it is subjected to the lowest possible pressure – say,
for instance, 46 mm.
If possible the apparatus should always be kept filled
with the thermal medium.
8. Experiments

8.1 Experiment set-up:

Additionally required:
1 Bath/circulating thermostat
1 Dig. quick-response pocket thermometer U11853
1 Type K NiCr-Ni immersion sensor,
-65°C-550°C
2 Silicone tubes, 1 m
1 l Anti-freeze fluid with corrosion-inhibiting additive
for aluminium engines (e.g., Glysantin® G30 ma-
nufactured by BASF)
Place the equipment at a suitable height so that it
is convenient to observe the measuring cell. Posi-
tion it so that the safety valve does not point in
the direction of any people who could be injured
or objects that could be damaged.
Connect the silicone tubing from the outlet of the
circulation thermostat to the inlet of the heat cas-
ing and from the outlet of the heat casing to the
inlet of the circulation thermostat.
Prepare the thermal medium consisting of 2 parts
water to 1 part anti-freeze by volume.
Fill the circulated thermostat bath.
8.2 Qualitative observations:
Liquid and gaseous states, dynamic state during phase
transformation, transition points occurring at differ-
ent temperatures.
Vary the volume by turning the handwheel and
the temperature by means of the thermostat. Ob-
serve the safety instructions while doing so.
Carefully shake the set-up to conduct simple
observations on the boundary between liquid and
gas.
In the vicinity of the critical point, it is also possible to
observe the critical opalescence. Owing to the con-
stant changing of state between liquid and gaseous
states in small regions of the measuring cell, a kind of
"mist" develops and the sulphur hexafluoride appears
to be turbid.
8.3 Measuring isotherms in a p-V diagram:
At maximum volume, set the desired temperature
on the circulation thermostat.
Gradually reduce the volume in the measuring
cell (in steps down to a position of 10 mm). Wait
till a stationary equilibrium has been attained be-
fore taking pressure readings.
Then, beginning with the minimum volume,
gradually increase the volume till the piston posi-
tion is once again at 10 mm. Wait till a stationary
equilibrium has been attained before taking pres-
sure readings.
Convert the excess pressure readings into abso-
lute pressure and the piston positions into vol-
ume, as described in chapter 6.
In the low-volume region, stationary equilibrium is
attained more quickly during transition from higher
to lower pressure – i.e. from a lower volume to a
greater volume – since the phase boundary layer for
the phase transition from liquid to gas is created by
vapour bubbles present throughout the liquid. Sta-
tionary equilibrium then takes around 1 to 5 minutes
U14400
to attain, whereby the measurements on the fringe of
the region where both phases exist take longest.
The recommended threshold value of 10 mm refers to
U11854
a filling pressure of 10 bar. Above this value, there
U10146
will certainly be no occurrence of a liquid phase in
the permissible temperature range. The threshold
value shifts to the "right" if the filling pressure is
higher.
8.4 Measuring isochores in a p-T diagram:
Set the desired initial temperature. Subsequently
set the desired volume.
Gradually allow the temperature to decrease.
Wait till a stationary equilibrium has been at-
tained then take the pressure reading.
Measurements where both phases are present can be
plotted to generate a vapour-pressure curve.
Attainment of equilibrium takes up to 20 minutes
after each change of temperature due to the fact that
7

Publicité

Table des Matières
loading

Table des Matières