3B SCIENTIFIC PHYSICS U104001 Instructions D'utilisation page 20

Appareil d'analyse du point critique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
Filling begins with several flush cycles in which the air
is flushed out of the pipe. The number of cycles re-
quired to flush out the air depends on the length of
the pipe (more precisely, on the ratio of the pipe
length to the volume of the measuring cell). In the
process, care should be taken that the quantity of the
greenhouse gas SF
released in the environment is
6
reduced to a minimum.
Connecting a fixed pipe:
Fig. 3: Connecting a fixed pipe
(a) flush valve, (b) regulating valve
If necessary, pull out the protection for the gas
connection and loosen the valve nut (11 mm) to
remove the 1/8" gas connection fittings.
Connect the pipe (if necessary with adapters) to
the gas fitting.
Beginning with the valve nut, slide the supplied
screw joints onto the tubing. (See Fig. 3: follow
the sequence and alignment specified along with
the cable binder)
Insert the pipe into the regulating valve and
tighten the valve nut till the point is reached
where it is no longer possible to move the pipe
any further using only your fingers.
Hold the regulating valve still with an open-end
spanner (13 mm) and tighten the valve nut by a
further 270°.
Now, the connection is gas-tight. When loosening the
valve nut afterwards, the regulating valve also needs
to be held still with a spanner.
Flushing out air:
Use the handwheel to set the piston position to
10 mm.
Slowly open the regulating valve and let in the SF
till a pressure of approx. 10 bar has been at-
tained.
Shut the regulating valve.
0
10
20
0 0
1 9
1 8
1 7
1 6
1 5
a
b
Open the flush valve slightly till the pressure has
dropped to almost 0 bar.
Shut the flush valve.
Filling with test gas:
After at least four flush cycles, open the regulat-
ing valve till the pressure attained is once again
10 bar.
Shut the regulating valve.
Turn the handwheel in the reverse direction till
the piston reaches a position of say 46 mm.
Slowly open the regulating valve and shut it again
when a pressure of 10 bar has been attained.
30
40
50mm
7.3 Filling with gas from a MINICAN®:
Additionally required:
1 MINICAN® gas container with SF
company Westfalen (www.westfalen-ag.de
If the equipment is used only occasionally, it is more
practical to draw the test gas from a MINICAN® gas
container. The gas connection of a MINICAN® con-
tainer is similar in design to a commercial spray can,
i.e. it opens when the MINICAN® container is pressed
directly onto the gas connection fittings.
Here too, filling begins with several rinsing cycles for
flushing out the air.
SF
6
Fig. 4: Filling with test gas from a MINICAN® gas container
(a) flush valve, (b) regulating valve
Flushing out air:
If necessary, pull off the protection for the gas
connection.
Use the handwheel to set the piston position to
10 mm.
After removing the protective cap, position the
MINICAN® container with SF
6
tion fittings.
Press the MINICAN® container onto the gas con-
nection fittings, slowly open regulating valve (b)
6
, e.g. from the
6
0
a
b
onto the gas connec-
6
10
20
30
40
50mm
0 0
1 9
1 8
1 7
1 6
1 5

Publicité

Table des Matières
loading

Table des Matières